Test Results
220
MAX-800 Series
Alarms/Errors Overview
Ethernet - PCS Lanes
Note:
Available with parallel interfaces only: 100 Gbit/s Ethernet Available under
the
PCS
sub tab for Through Mode test providing alarms/errors for both
ports.
Skew Alarm Threshold (bits)
, available with Ethernet test application,
allows setting the threshold value that will be used to declare a skew
alarm. Not available at 100G when the RS-FEC check box is selected.
Default
restore the default skew alarm threshold value.
Alarms
Note:
For
LOA
and
Inv. Mapping
alarms see
LOBL
(Loss of Block Lock)
RX
: In lock mode, 65 invalid sync fields (00 or 11) within a 1024 sync
window are received. A PCS Lane Lock alarm is cleared when
receiving 64 consecutive valid 66b sync field (01 or 10) in a row.
LOAML
(Loss of Alignment Marker Lock)
RX
: In lock mode, four consecutive marker values are received that do
not match the alignment marker that the lane is currently locked to.
The LOAML alarm is cleared when the PCS Lane is declared Lock and
two valid alignment markers 16384 blocks (66b) apart are received.
Rate
Range
Default Value
100G
0 to 2047
928
Summary of Contents for MAX-800 Series
Page 1: ...www EXFO com User Guide MAX 800 Series MaxTester Application ETHERNET AND TRANSPORT TESTER...
Page 16: ......
Page 18: ......
Page 26: ......
Page 43: ...Test Setup Test Applications MaxTester 35 SONET SDH BERT For SDH BERT...
Page 48: ...Test Setup Test Applications 40 MAX 800 Series SONET SDH DSn PDH BERT For SDH...
Page 62: ......
Page 208: ......
Page 414: ......
Page 426: ......
Page 473: ...Glossary MaxTester 465 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values...
Page 476: ......
Page 494: ...0 5 1 5 48 5 0 176 W d y K y K D...