Test Functions
MaxTester
363
Pointer Adjustment
Jump
New Pointer
allows selecting a new pointer value:
For STS/AU:
0
(default) to
782
For VT/TU:
Inject
button sends the new pointer value.
New Data Flag
(NDF) check box when selected inserts a New Data
Flag with the pointer adjustment when the Inject button is tapped.
For STS/AU: When NDF is enabled, bits 1 to 4 of the pointer word (H1
and H2 bytes) are set to
1001
when executing a pointer jump.
For VT/TU: When NDF is enabled, bits 1 to 4 of the pointer word (V1
and V2 bytes) are set to
1001
when executing a pointer jump.
Path
Range
VT1.5
0 to 103
VT2
0 to 139
TU-3
0 to 764
TU-11
0 to 103
TU-12
0 to 139