Test Setup - Test Configurator, Timer, and System
108
MAX-800 Series
FTFL/PT
FTFL/PT
From the
Test
menu, tap
Setup
,
Test Configurator
, tap on the signal block,
and on the
FTFL/PT
tab.
FTFL
Allows the configuration of the Forward and Backward ODU Fault Type
Fault Location (FTFL) to be generated.
Fault Indication
and
Fault Indication Code
allow the selection of the
FTFL fault indicator message/code (byte 0 for forward, byte 128 for
backward) to be generated.
Note:
The
Fault Indication Code
field is automatically updated when the
Fault
Indication
is changed and vice versa.
Operator Identifier
allows editing the Operator Identifier to be
generated (bytes 1 to 9 for forward, bytes 129 to 137 for backward;
9 characters allowed). By default no Operator Identifier is defined.
Operator Specific
allows editing the Operator Specific to be
generated (bytes 10 to 127 for forward, bytes 138 to 255 for backward;
118 characters allowed). By default no Operator Specific is defined.
Fault Indication
Fault Indication Code (Hex)
No fault
00
(default)
Signal fail
01
Signal Degrade
02
Reserved
03
a
a.
Selecting
Reserved
will use the hexadecimal code
03
but, all codes from
03
to
FF
are reserved for future international standardization.
Summary of Contents for MAX-800 Series
Page 1: ...www EXFO com User Guide MAX 800 Series MaxTester Application ETHERNET AND TRANSPORT TESTER...
Page 16: ......
Page 18: ......
Page 26: ......
Page 43: ...Test Setup Test Applications MaxTester 35 SONET SDH BERT For SDH BERT...
Page 48: ...Test Setup Test Applications 40 MAX 800 Series SONET SDH DSn PDH BERT For SDH...
Page 62: ......
Page 208: ......
Page 414: ......
Page 426: ......
Page 473: ...Glossary MaxTester 465 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values...
Page 476: ......
Page 494: ...0 5 1 5 48 5 0 176 W d y K y K D...