Operation
Liquiline System CA80PH
74
Hauser
Overflow warnings
Analyzer event logbook="Fill
up buffer"
Options
• Off
• On
Factory setting
Off
If the fill-up buffer overruns, you can decide whether
you want to receive a diagnostic message from the cont-
roller or not for each individual logbook.
Analyzer data logbooks
For analyzer measured data
The entry is made automatically on completion of the
measurement. A setting does not need to be made. The
logbook is activated automatically. SP1 is assigned to
data logbook SP1.
Data logbook SP1
Assignment to the measuring channel
Source of data
Read only
The assigned measuring channel is displayed
Measuring parameter
Read only
Plain-text information on the parameter that is being
recorded
Main value
Read only
Information on the main value and the unit.
Unit
Read only
Logbook name
Customized text,
16 characters
Line plotter
Menu to define the graphic display
Axes
Options
• Off
• On
Factory setting
On
Should the axes (x, y) be displayed (On) or not (Off)?
Orientation
Options
• Horizontal
• Vertical
Factory setting
Horizontal
You can choose whether the value curves should be
displayed from left to right ("Horizontal") or from top to
bottom ("Vertical").
If you want to display two data logbooks simultaneously,
make sure that both logbooks have the same settings
here.
X-Description
Options
• Off
• On
Factory setting
On
Decide whether a description should be displayed for the
axes and whether grids should be shown. In addition,
you can also decide whether a pitch should be displayed.
Y-Description
Grids
Pitches
X Pitch/Grid distance 10 ... 50%
Factory setting
10 %
Specify the pitch.
Y Pitch/Grid distance
Path: Menu/Setup/General settings/Logbooks
Function
Options
Info