Liquiline System CA80PH
Operation
Hauser
89
10.2.3 Calibration
Path: Menu/Setup/Analyzer/Calibration
Function
Options
Info
Start condition
Options
• Immediate
• Date/time
Factory setting
Immediate
The calibration can start either immediately or on the
set date/time.
If Start condition
Date/time
is selected:
Date
01.01.1970 ... 07.02.2106
Factory setting
DD.MM.YYYY
Time
00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24 h)
Calibration interval
00-01 to 90-00 (DD-HH)
Factory setting
02-00
For configuring the time interval of the calibra-
tion/adjustment.
Next calibration
Mode="Automatic"
Read only
Calibration factor
Read only
Relationship of the measured concentration to the
pre-specified concentration of the calibration standard.
Zero point
Read only
Settings
Nominal concentration
1.0 to 10.00 (mg/l)
Factory setting
For configuring the concentration of the calibration
standard solution.
Depends on the settings under Menu/Setup/Analy-
zer/Extended setup/Measuring value/Main value
Automatic cleaning
Options
• Off
• On
Factory setting
On
For specifying whether cleaning will take place before
every calibration/adjustment (only in the automatic
mode).