Electrical connection
Liquiline System CA80PH
44
Hauser
LEDs on front of module
LED
Name
Color
Description
RJ45
LNK/ACT
GN
• Off = Connection is not active
• On = Connection is active
• Flashing = Data transmission
RJ45
10/100
YE
• Off = Transmission rate 10 MBit/s
• On = Transmission rate 100 MBit/s
PWR
Power
GN
Supply voltage is applied and module is initialized
BF
Bus failure
RD
not used
SF
System failure
RD
System failure
COM
Communication
YE
Sending or receiving Modbus message
DIP switches on front of module
DIP
Factory setting
Assignment
1-128
ON
not used
OFF
Write protection: "ON" = configuration is
not
possible via Ethernet, only via local operation
Service OFF
If the switch is set to "
ON"
, the user settings for Ethernet addressing are saved and
connection settings programmed into the device at the factory are activated:
IP address=192.168.1.212, subnet mask=255.255.255.0, gateway=0.0.0.0, DHCP=Off
If the switch is set to "
OFF"
, the saved user settings are activated again.
Summary of Contents for Liquiline System CA80PH
Page 2: ......
Page 208: ...www addresses endress com ...