Operation
Liquiline System CA80PH
76
Hauser
Source of data
Read only
This is for information purposes only. If you want to
record another value, delete this logbook and create a
new data logbook.
Measured value
Log time left
Data logbook="Fill up
buffer"
Read only
Displays the days, hours and minutes remaining until
the logbook is full.
Log size
Data logbook="Ring buffer"
Read only
Displays the number of entries remaining until the
logbook is full.
Logbook name
Customized text,
20 characters
You can change the name here again.
Scan time
00:00:01 ... 01:00:00
Factory setting
00:01:00
As above
Minimum interval between two entries
Format: HH:MM:SS
Must be configured individually for each data logbook
Data logbook
Options
• Off
• Ring buffer
• Fill up buffer
Factory setting
Off
Ring buffer
If the memory is full, the latest entry automatically
overwrites the oldest entry.
Fill up buffer
If the memory is full, there is an overflow, i.e. no new
values can be saved. The controller displays a
corresponding diagnostic message. The memory then
has to be cleared manually.
Line plotter
Menu to define the graphic display
Axes
Options
• Off
• On
Factory setting
On
Should the axes (x, y) be displayed (On) or not (Off)?
Orientation
Options
• Horizontal
• Vertical
Factory setting
Horizontal
You can choose whether the value curves should be
displayed from left to right ("Horizontal") or from top to
bottom ("Vertical").
If you want to display two data logbooks simultaneously,
make sure that both logbooks have the same settings
here.
X-Description
Options
• Off
• On
Factory setting
On
Decide whether a description should be displayed for the
axes and whether grids should be shown. In addition,
you can also decide whether a pitch should be displayed.
Y-Description
Grids
Pitches
X Pitch/Grid distance
10 ... 50%
Factory setting
10 %
Specify the pitch.
Y Pitch/Grid distance
Remove
Action
This action removes the data logbook. Any data that
have not been saved are lost.
Path: Menu/Setup/General settings/Logbooks
Function
Options
Info