Liquiline System CA80PH
Diagnostics and troubleshooting
Hauser
145
285 Update error
F
On
On
Firmware update failed
Possible reasons:
• Read error
• SD card is faulty
• Incorrect firmware on SD card
1. Repeat update
2. Use another data storage medium with
firmware
3. Check data storage medium for correct
firmware
302 Battery low
M
On
Off
Buffer battery of real time clock is low
It is located on the backplane. The date and time
are lost if the power is interrupted.
• Replace battery
304 Module data
F
On
On
At least 1 module has incomplete configuration
data
• Check module information
305 Power consumption
F
On
On
Total power consumption is too high
1. Check installation
2. Remove sensors/modules
3. Restart the device
306 Software error
F
On
On
Internal firmware error
Device configuration does not match firmware
• Contact the Service Department
312 Temperature sensor
F
On
On
Temperature sensor PT3 for ambient
temperature measurement is faulty.
• Winter operation regulation not possible
• Sampling and distribution arm blocked to
protect against freezing
1. Deactivate winter operation mode at
Setup/Inputs/Temperature S:3/Winter
operation
2. Replace sensor
360 Cooling/Heating
S
On
Off
Temperature range in the housing exceeded
1. Check installation conditions and the
ambient temperature
2. Replace the FMAB1 module
361 Cooling/Heating
F
On
On
Cooling/heating module defective
Defined temperature range not reached. This
could affect the functionality of the reagents.
1. Check whether the insulation cap is seated
correctly over the reagents
2. Replace cooling/heating module
362 Photometer temp.
F
On
Off
Temperature of the photometer too high
1. Replace photometer
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current