Liquiline System CA80PH
Diagnostics and troubleshooting
Hauser
147
380 Firmware failure
F
On
On
Internal software error
1. Update the software
2. Replace backplane
3. Contact the service department and quote
the number displayed
401 Factory reset
F
On
On
Factory reset is performed
405 Service IP active
C
On
Off
Service switch is switched on.
The device can be addressed at 192.168.1.212.
Switch off the service switch to change to the
saved IP settings.
--> disable the service IP
406 Param. active
C
Off
Off
Configuration enabled
--> Wait for configuration to be finished
407 Diag. active
C
Off
Off
Diagnostic menu enabled
--> Wait for maintenance to be finished
408 Calibration aborted
M
Off
Off
Calibration was aborted
412 Writing backup
F
On
Off
--> Wait for the write process to be finished
413 Reading backup
F
On
Off
--> Wait
460 Output below limit
S
On
Off
Current output not reached
Measured value is outside the specified current
range
Possible reasons:
• Sensor in air
• Air pockets in assembly
• Incorrect flow to sensor
• Sensor fouled
Measures
1. Check the application
2. Check configuration of current output
3. Clean sensor
461 Output above limit
S
On
Off
Current output exceeded
Measured value is outside the specified current
range
Possible reasons:
• Sensor in air
• Air pockets in assembly
• Incorrect flow to sensor
• Sensor fouled
Measures
1. Check the application
2. Check configuration of current output
3. Clean sensor
502 No text catalog
F
On
On
No text catalog installed
--> Contact the Service Department
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current