Operation
Liquiline System CA80PH
94
Hauser
If Installation type
Pressurized pipe sampling
is selected (with Liquiline System CAT810, for example):
Sample preparation 1
In the case of a two-channel device, Sample preparation
2 is also displayed.
Cleaning valve
Options
• Not available
• Available
Factory setting
Not available
In the event of retrofitting, the valve can be activated
here.
If Cleaning valve
Available
is selected:
Filter backflush
Options
• On
• Off
Factory setting
On
If Filter backflush
On
is selected:
Cleaning interval
00:10 to 02:00 HH:MM
Factory setting
00:30
Compressed air or water cleaning (optional) for longer
filter maintenance intervals.
Cleaning duration
10 to 30 s
Factory setting
10 s
Cleaning duration of the automatic water or compressed
air cleaning system
Cleaning discard time
0 to 1800 s
Factory setting
180 s
Discard time once cleaning is complete. If water is used
as the flushing medium, this must be replaced by a fresh
sample before starting the next measurement.
Reset settings
Resets all settings specific to sample preparation All
other settings are retained.
Path: Menu/Setup/Sample preparation
Function
Options
Info