Liquiline System CA80PH
Operation
Hauser
91
10.3 Sample preparation
The menu displayed depends on the connected sample preparation system. The menu item
does not appear if a Liquiline System CA80PH analyzer is connected without a collecting
vessel.
The Liquiline System CAT860 can only be operated with a Liquiline System CA80
single-channel device.
Path: Menu/Setup/Sample preparation
Function
Options
Info
Used channels
Options
• SP1
• SP2
• SP1 + SP2
Factory setting
SP1
Is only displayed for a two-channel device.
For further information, see the "Measurement" section.
Installation type
Options
• Others
• Pressurized pipe
sampling
Factory setting
Others
Others
:
e.g. Liquiline System CAT820 / CAT860
Pressurized pipe sampling
:
Sample preparation that is installed in a pressure pipe,
e.g. Liquiline System CAT810.
If Installation type
Others
is selected:
Sample preparation 1
In the case of a two-channel device, Sample preparation
2 is also displayed.
Operating mode
Options
• Controlled
• Independent
Factory setting
Controlled
Controlled:
Sample preparation that is controlled by analyzer CA80,
e.g. Liquiline System CAT820 / CAT860
Independent:
No sample preparation that is controlled by the Liquiline
System CA80 analyzer, e.g. Stamoclean CAT430.
The supply of sample must be guaranteed by the custo-
mer.
If Operating mode
Controlled
is selected:
Tag
Customized text,
32 characters
Cleaning start condition
(CAT860)
Options
• Immediate
• Date/time
• Disabled
Immediate:
Cleaning starts immediately.
Date/time:
Cleaning starts at the selected date/time.
Disabled:
Cleaning is not active.
If Start condition
Immediate
or
Date/time
is selected:
Cleaning interval
(CAT860)
00-01 to 90-00 (DD-HH)
Factory setting
01-00
Compressed air or water cleaning for longer filter
maintenance intervals
Cleaning with liquid cleaner