System integration
Liquiline System CA80PH
48
Hauser
7
System integration
7.1
Web server
7.1.1
Connection
‣
Connect the PC communication cable to the RJ45 port of module 485 or module ETH.
a0016228
Fig. 61: Ethernet connection
7.1.2
Creating the data connection
1. Start your PC.
2. First, set a manual IP address in the network connection settings of the operating system.
This address must be in the same subnetwork as the IP address of the device.
Example:
• IP address for Liquiline System: 192.168.1.212
(Diagnose/Systeminformationen/Ethernet/IP-Adresse)
• IP address for PC: 192.168.1.
213
3. Start the Internet browser.
If you use a proxy server to connect to the Internet:
4. Disable the proxy (browser settings under "Connections/LAN settings").
5. Enter the IP address of your device in the address line.
The system takes a few moments to establish the connection and then the CM44 web
server starts.