Diagnostics and troubleshooting
Liquiline System CA80PH
146
Hauser
363 Photometer temp.
F
On
Off
Temperature of the photometer too low
1. Replace photometer
364 Timeout dosing
F
On
On
The liquid manager/dosing dispenser has timed
out.
Possible reasons:
• Photoelectric barrier defective
• Blocking
-> Contact the Service Department
365 Photometer comm.
F
On
On
Photometer not communicating
Possible reasons:
• Incorrect photometer connection
1. Check photometer connection
366 Dosing unit defect.
F
On
On
Dosing unit is faulty
-> Contact the Service Department
367 Module connect.
F
On
On
No communication to sample preparation
-> Check connecting cable to sample preparation
system.
368 Reactor temp. is low
F
On
On
Target temperature of decomposition reactor not
reached
-> Contact the Service Department
369 Reactor temp. is high
F
On
On
Temperature of decomposition reactor is too high
1. Check the reactor temperature on the
display. If the reactor temperature does not
drop, disconnect the power supply.
2. Contact the Service Department
370 Internal voltage
F
On
On
Internal 24-V voltage outside the valid range
--> Check supply voltage
372 Dosing failure
F
On
On
Error in dosing process
1. Check hoses and connectors
2. Check level
3. Replace dosing dispenser
4. Contact the Service Department
373 Electronictemp. high
M
On
Off
Temperature of backplane electronics is high
--> Check ambient temperature and energy
consumption
374 Sensor check
F
On
Off
Sensor measuring sequence counter
No measurement signal from sensor
1. Check sensor, replace if necessary
2. Check sensor connection
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current