Operation
Liquiline System CA80PH
112
Hauser
10.6.3 Modbus RS485 and Modbus TCP
Specify which process values should be output via Modbus RS485 communication or via Modbus
TCP.
In the case of Modbus RS485, you can switch between the RTU and the ASCII protocol.
You can define a maximum of 16 device variables:
‣
Define the data source. You can choose from the analyzer as well as sensor inputs and
controllers.
‣
From the source, select the measured value that should be output.
‣
Define how the device should behave in the event of a "Hold" (for sensors).
‣
The configuration options for "Source of data", "Measured value" and "Hold behavior" are
identical to those described in the "Outputs/Current outputs" section. Please refer to this
section for more information.
‣
Please note that if you select Hold behavior="Freeze", the system not only flags the status but
also actually "freezes" the measured value.
Additional information is provided in the SD01189C document.
10.6.4 EtherNet/IP
Determine which process values should be output via EtherNet/IP communication.
You can define a maximum of 16 analog device variables (AI):
‣
Define the data source. You can choose from the analyzer, as well as sensor inputs, binary
inputs and controllers.
‣
From the source, select the measured value that should be output.
‣
Define how the device should behave in the event of a "Hold" (for sensors).
‣
For controllers, you also specify the manipulated variable type.
‣
Please note that if you select Hold behavior="Freeze", the system not only flags the status but
also actually "freezes" the measured value.
In addition, you can define 8 digital device variables (DI):
‣
You specify only the data source here. You can choose from relays, binary inputs and limit
switches.
Additional information can be found on the product pages on the internet (SD01293C).