Electrical connection
Liquiline System CA80PH
30
Hauser
6.2.3
Connecting 24V devices
• To access the electronics compartment, proceed as described in the "Routing the cables"
section.
• Guide the 24V connecting cable from below through the cable gland on the inner rear panel
of the device and feed it upwards into the electronics compartment.
• Establish the connection as per
a0023128
Fig. 22: Terminal assignment
1
Assignment: 100 to 120 V/200 to 240 V AC ±10 %
2
Assignment: 24 V DC +15/-9 %
3
Internal 24 V voltage
L
N
N
L
-
-
+
+
1
2
3
Summary of Contents for Liquiline System CA80PH
Page 2: ......
Page 208: ...www addresses endress com ...