Data management
Firmware update
Please contact your local sales office for information on firmware updates available for your
controller and its compatibility with earlier versions.
The
current firmware version
of the analyzer, FXAB1 control module, photometer and sample
preparation 1 can be found at: Menu/Diagnostics/System information/Software version.
First save your current setup on an SD card since a firmware update overwrites your
settings with the factory settings. After updating the firmware, you can restore your setup
by uploading it from the SD card.
To install a firmware update, you must have the update available on an SD card.
1. Insert the SD card into the controller card reader.
2. Go to: Menu/Setup/General settings/Extended setup/Data management/Firmware update.
The update files on the SD card are displayed.
3. Select the desired update and select yes when the following question is displayed: The
current firmware will be overwritten. After this the device will reboot. Do you want to
proceed?
The firmware is loaded and the device is then started with the new firmware.