Liquiline System CA80PH
Electrical connection
Hauser
33
6.3.2
Liquiline System CAT820 / CAT860 connection
Only use terminated genuine cables. The cables must be shielded.
1. Disconnect the mains plug.
2. Remove the bottle tray by lifting up the recessed grip slightly and then pulling it towards the
front.
3. Release a suitable hose gland on the right-hand underside of the analyzer and remove the
dummy plug from the gland.
4. Guide the spiral hose through the hose gland.
5. Remove the protective cover in the upper right-hand corner.
Cable example (does not necessarily correspond to the original cable)
a0013210
Fig. 30: Terminated cable
1
Outer shield (exposed)
2
Cable cores with ferrules
3
Cable sheath (insulation)
a0023132
Fig. 31: Inserting the cable
4
Grounding clamp for Memosens
supply and power supply
5
Cable clamp for hose heating
a0023133
Fig. 32: Tightening the screw (2 Nm)
1)
The cable shield is grounded by the
grounding clip
1)
Observe the information in the "Guaranteeing the degree of protection" section (
1
2
3
SP1
N
L
PE
86
85
97
98
4
5
SP1
N
L
PE
86
85
97
98