Liquiline System CA80PH
Operation
Hauser
83
Loading the setup
You can load a setup you have saved quickly and easily
When you load a setup, the current configuration is overwritten. Note that cleaning and
controller programs could be active. Do you want to continue anyway?
1.Insert the SD card into the controller card reader.
2. Go to: Menu/Setup/General settings/Extended setup/Data management/Load setup.
A list of all the setups on the SD card is displayed.
3. Select the desired setup.
The device then displays the following message:The current parameters will be
overwritten and the device will reboot. Warning: Please note that cleaning and
controller programs can be active. Do you want to proceed?
4. Select "OK" to confirm or cancel the action.
The desired setup is restored after restarting the device.
Exporting the setup
Exporting the setup gives you the following advantages
• Export in XML format with a stylesheet for formatted display in an XML-compatible
application such as Microsoft Internet Explorer
• Importing the data (drag and drop the XML file into a browser window)
1. Insert the SD card into the controller card reader.
2. Go to: Menu/Setup/General settings/Extended setup/Data management/Export setup.
3. Assign a file name (Name).
4. Then select "Export".
5. If you have already assigned the file name, you will be asked whether you want to overwrite
the existing setup.
6. Select "OK" to confirm, or cancel the action and give the file a new name.
Your setup is saved on the SD card in a folder named "Device".
You cannot upload the exported setup to the device again. You must use the "Save setup"
function for this purpose. This is the only way you can save a setup to an SD card and reload
it later on or upload it to other devices.
Summary of Contents for Liquiline System CA80PH
Page 2: ......
Page 208: ...www addresses endress com ...