10.2.4 Cleaning
Additional cleaning can be adapted to the application conditions in this menu.
Path: Menu/Setup/Analyzer/Cleaning
Function
Options
Info
Start condition
Options
• Immediate
• Date/time
• Disabled
Factory setting
Disabled
Immediate:
Cleaning of the valve manifold and cell starts
immediately.
Date/time:
Cleaning of the valve manifold and cell starts on the
selected date/time.
Disabled:
Additional cleaning is not enabled.
If Start condition
Date/time
is selected:
Date
01.01.1970 ... 07.02.2106
Factory setting
DD.MM.YYYY
Time
00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24 h)
Cleaning interval
00-01 to 90-00 (DD-HH)
Factory setting
02-00
For configuring the time interval for (additional) clea-
ning.
Next cleaning
Mode="Automatic"