Diagnostics and troubleshooting
Liquiline System CA80PH
142
Hauser
–
M
= Maintenance required: Action must be taken as soon as possible.
The device stills measures correctly. Immediate measures must not be taken. However,
proper maintenance efforts would prevent a possible malfunction in the future.
• Message text
If you contact the Service Department, please cite the message number only. Since you can
individually change the assignment of an error to an error category, the Service Department
cannot use this information.
11.5.2 Adapting the diagnostic behavior
All the diagnostics messages are assigned to specific error categories at the factory. Since other
settings might be preferred depending on the application, error categories and the effect errors
have on the measuring point can be configured individually. Furthermore, every diagnostics
message can be disabled.
Example
The controller returns diagnostics message 531 "Logbook full". You want to change this message
so that an error is not indicated on the display for example.
1.
Go to:
– Menu/Setup/General settings/Extended setup/Diagnostics/Device behavior
for device-specific diagnostics messages (as in this example)
– Menu/Setup/Analyzer/Extended setup/Diagnostics settings
for analyzer-specific diagnostic messages.
– Menu/Setup/Inputs/<Sensor type>/Extended setup/Diagnostics settings/Diag.
behavior
for sensor-specific diagnostics messages.
2.
Select the diagnostics message and press the navigator button.
3.
Decide:
a. Should the message be deactivated?
b. Do you want to change the error category?
c. Should an error current be output?
d. Do you want to trigger a cleaning program?
4.
Deactivate the message, for example (Diagnostics message to "Off").