Liquiline System CA80PH
Electrical connection
Hauser
43
6.7.2
Module ETH
DIP switches on front of module
DIP
Factory setting
Assignment
1-128
ON
Bus address (--> "Commissioning/Communication")
OFF
Write protection: "ON" = configuration
not
possible via the bus, only via local operation
Service OFF
If the switch is set to "
ON"
, the user settings for Ethernet addressing are saved and
connection settings programmed into the device at the factory are activated:
IP address=192.168.1.212, subnet mask=255.255.255.0, gateway=0.0.0.0, DHCP=Off
If the switch is set to "
OFF"
, the saved user settings are activated again.
a0025316
Fig. 55: Module ETH
a0025317
Fig. 56: Wiring diagram for module ETH