Liquiline System CA80PH
Diagnostics and troubleshooting
Hauser
149
716 Calibration expired
S
On
Off
Calibration has failed or is not reliable
Possible reasons:
• Stability criteria are not met
-> Check the configuration and perform a manual
calibration again
-> Contact the Service Department
717 Photometer LED
F
On
On
Photometer defective
Possible reasons:
• No LED voltage
• No LED current
-> Contact the Service Department
718 Cuvette covering
M
On
Off
Photometer test
High level of contamination - reliable
measurement will not be possible soon
1. Clean analyzer with manual cleaning process
2. Contact the Service Department
3. Replace photometer
719 Cuvette covering
F
On
Off
Photometer test
Buildup, high level of pollution - measurement no
longer possible
1. Start cleaning
2. Contact the Service Department
726 Liquids warning
M
On
Off
Consumable liquids, warning
Level monitoring or shelf life
Measurement can still take place.
Possible reasons:
• The level of one or more liquids is low or the
expiry date of one or more liquids is
approaching.
Top up/replace liquids in question and reset
counter in Diagnostics/Operating time
information
727 Liquids alarm
F
On
Off
Consumable liquids, alarm
Level monitoring or shelf life
Measurement can still take place.
Possible reasons:
• The level of one or more liquids is low or one
or more liquids have exceeded their expiry
date.
Top up/replace liquids in question and reset
counter in Diagnostics/Operating time
information
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current