Liquiline System CA80PH
Operation
Hauser
77
Example for setting up a new data logbook (for sensors, for instance)
1.
Menu/Setup/General settings/Logbooks/Data logbooks/New:
a. Logbook name: Assign a name, e.g. "01".
b. Source of data: Select a data source, e.g. the sensor connected to channel 1 (CH1).
c. Measured value: Select the measured value that you wish to record.
d. Scan time: Specify the time interval between two logbook entries.
e. Data logbook: Activate the logbook. Specify the type of memory to be used here, "Ring
buffer" or "Fill up buffer".
2.
../Finished: Perform this operation.
--> Your new logbook now appears in the list of data logbooks.
3.
Select the data logbook with the name "01".
4.
If you selected "Fill up buffer", you can also choose whether you want to receive a diagnostic
message in the event of memory overrun.
5.
Depending on the type of memory selected, you receive information about the memory
space (for "Ring buffer") or the time remaining until memory overrun (for "Fill up buffer").
6.
Define the graphic display mode in the "Line plotter" submenu..
10.1.5 Extended setup
Diagnostics settings
The list of diagnostic messages displayed depends on the path selected. There are
device-specific messages, and messages that depend on what sensor is connected.
Path: ... /Extended setup/Diagnostics settings/Diag. behavior
Function
Options
Info
List of diagnostic messages
Select the message to be changed. Only then can you
make the settings for this message.
Diag. code
Read only
Diagnostic message
Options
• On
• Off
Factory setting
Depends on the message
You can deactivate or reactivate a diagnostic message
here.
Deactivating means:
• No error message in the measuring mode
• No error current at the current output
Error current
Options
• On
• Off
Factory setting
Depends on the message
Decide whether an error current should be output at the
current output if the diagnostic message display is
activated.
Status signal
Options
• Maintenance (M)
• Out of specification (S)
• Function check (C)
• Failure (F)
Factory setting
Depends on the message
The messages are divided into different error categories
in accordance with NAMUR NE 107.
Decide whether you want to change a status signal
assignment for your application.