Liquiline System CA80PH
Technical data
Hauser
199
15.7 Performance characteristics
15.7.1 Measured error
3)
15.7.2 Maximum measured error Sensor inputs
--> Documentation of the connected sensor
15.7.3 Measured error current inputs and outputs
Typical measured errors:
< 20
μ
A (for current values < 4 mA)
< 50
μ
A (with current values 4 to 20 mA)
at 25 °C (77° F) in each case
Additional measured error depending on the temperature:
< 1.5
μ
A/K
15.7.4 Repeatability
3)
• Blue method: ± 2% of the display value + 0.01 mg/l
• Yellow method: ± 2% of the display value + 0.05 mg/l
15.7.5 Repeatability Sensor inputs
--> Documentation of the connected sensor
15.7.6 Measuring interval
• Blue method: continuous (approx. 11 min), adjustable > 11 min
• Yellow method: continuous (approx. 8 min), adjustable > 10 min
15.7.7 Sample requirement
22 ml/measurement
3)
following ISO 15839 with standard solutions. Measured errors include all the uncertainties of the analyzer. They do
not include all the uncertainties from the standard solutions used as a reference.
CA80PH-AAE1:
0.05 to 2.5 mg/l PO
4
-P
+/- 2 % of end of measuring range
CA80PH-AAE2:
0.05 to 10 mg/l PO
4
-P
+/- 2 % of end of measuring range
CA80PH-AAE3:
0.5 to 20 mg/l PO
4
-P
+/- 2 % of end of measuring range
CA80PH-AAE4:
0.5 to 50 mg/l PO
4
-P
+/- 2 % of end of measuring range