Service
14.4 Maintenance
14.5 Repair
Repairs may only be performed by specialized service personnel.
It is recommended that only original parts from Dräger be used for repairs and that
the repairs be performed by Dräger.
The replacement of the integrated patient-gas measurement module is described in
the chapter "Replacing the patient-gas measurement module (PGM)" (see
"Replacing the patient-gas measurement module (PGM)", page 196).
Component
Interval
Measure
User group
CO
2
absorber / soda
lime with dust filter
When colored violet or
according to the config-
ured Infinity ID replace-
ment interval
Replace, see page 75
Users
Water trap (option)
As required, when
soiled, or according to
the configured Infinity ID
replacement interval
Replace, see page 77
Users
Flow sensors
As required, if calibra-
tion is no longer possi-
ble, or according to the
configured Infinity ID
replacement interval
Replace, see page 198 Users
Piston diaphragm
Annually
Replace, see page 236 Users
Inspiratory O
2
sensor
(option)
Every 2 years
Replace, see page 55
Users
AGS
As required, when
soiled
Replace
Specialized service per-
sonnel
AGS filter
If required, if filter is
soiled, or flow is no lon-
ger achieved
Replace
Users
O-rings on the water
trap mount (option)
Every 2 years
Replace
Service personnel
Pressure reducer
(option)
Every 6 years
Replace
Specialized service per-
sonnel
Filter cloth
– Above the gas mixer
– Power supply unit
Every 2 years
Replace
Service personnel
CLIC adapter (option)
Every 4 years
Replace
Users
Lead-gel battery
(2 pieces)
Every 2 years
Replace
Service personnel
Or after determination of
remaining capacity
Replace
Specialized service per-
sonnel