![Dräger Atlan A300 Instructions For Use Manual Download Page 241](http://html.mh-extra.com/html/drager/atlan-a300/atlan-a300_instructions-for-use-manual_2526728241.webp)
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
241
Reprocessing
13.9.2.4 APL
valve
►
WARNING
Risk of damage to breathing system
If the APL valve is not removed before the breathing system is reprocessed, it
may be damaged by the reprocessing measures. This can lead to leaks in the
breathing system.
►
Always remove the APL valve prior to reprocessing.
Using the lower knurled nut (1), unscrew the APL valve from the upper part of
the breathing system.
13.9.2.5
Non-return valves and fresh-gas decoupling valve
On the top side of the lower part of the breathing system:
►
Remove the yellow inspiratory valve (1) and the blue expiratory valve (2) from
the lower part of the breathing system.
39498
37058