Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
171
Configuration
11.3.3.5
Vertical tab "Config. 2"
11.3.3.6
Vertical tab "Config. 3"
Headline/
Parameter
Setting range
Description
Deactivate the alarm limit in the following modes: Man/Spon,
Pause, CPAP/PSV with
∆
Psupp < 5, ext. FGO
FiO
2
high
Yes;
No
Yes;
No
Yes;
No
Specifies the alarm behavior when
changing to a different ventilation
mode.
These settings apply only to a change
to a ventilation mode with lower or no
breathing support (see page 299).
The alarm behavior at the start of the
therapy is defined by the configuration
in the vertical tab
Limits
.
MV low
Yes
; No
MV high
Yes
; No
xMAC low
Yes;
No
etCO
2
low
Yes
; No
etCO
2
high
Yes
; No
inCO
2
high
Yes
; No
inAgent low
Yes
; No
Deactivate the alarm limit in cardiac bypass mode (CBM)?
FiO
2
high
Yes;
No
Specifies the alarm behavior in the
CBM mode.
MV low
Yes
; No
MV high
Yes
; No
inAgent low
Yes
; No
Headline/
Parameter
Setting range
Description
Alarm limits for "Cylinder almost empty"
O
2
[bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
Specifies the alarm limits for the supply
pressure of connected gas cylinders.
Air
[bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
N
2
O
[bar]; [kPax100]: Off; 15 to 40
20
[psi]: Off; 218 to 580
290