Tests
The checklist can be presented in tabular form or as a walk-through mode.
In the system setup, it is possible to specify whether tests always start in the walk-
through mode, see page 177.
On devices with an O
2
sensor, the system test checks whether an O
2
calibration is
required. If more than 7 days have passed since the last calibration of the O
2
sensor, an O
2
calibration will be automatically performed.
On devices with an integrated patient-gas measurement module, calibrations are
performed regularly during operation.
1.
WARNING
Risk of patient injury
During the system test, the device is pressurized.
►
To prevent patient injury, do not perform the system test and leakage test if a
patient is connected.
Touch the button for the system test.
2. Complete the tabular checklist (1).
It is possible to switch from the tabular checklist to the walk-through mode, if
required. To do this, touch the button (2). For further information see: "Checklist
in the walk-through mode", page 136.
3. If the O
2
sensor is to be calibrated despite a valid calibration, touch the
Calibrate
button (3).
4.
CAUTION
Risk of device malfunction and/or patient injury
Existing malfunctions cannot be detected if the system test is canceled. This
may result in device malfunctions and the patient may be put at risk.
►
More attention is required when operating without a system test.
►
Perform the system test every day. If the system test is canceled during
execution, perform it again as soon as possible.
3507
2
1
2
5
4
3
System test