![Dräger Atlan A300 Instructions For Use Manual Download Page 39](http://html.mh-extra.com/html/drager/atlan-a300/atlan-a300_instructions-for-use-manual_2526728039.webp)
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
39
Overview
Backup manual mode with a mechanically controlled gas mixer
1. Open the flap (1).
2. Activate the backup manual switch (2). Follow the instructions on the product
label (3) or the screen (4).
3. Ventilate the patient manually.
a. During mechanical ventilation or in the
Man/Spon
ventilation mode:
Ventilate manually with the breathing bag.
b. When using the external fresh-gas outlet:
Ventilate manually with the breathing bag on the non-rebreathing system.
4. Set the fresh-gas flow.
5. Check the vaporizer setting.
40677
39763