Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
263
Service
14.3.1 Remote
service
The device supports the following remote service functions:
– Help Ticket
– Remote Device Check
– Proactive call
Further information on the remote service function can be obtained from Dräger.
14.3.2 Safety
checks
Safety checks are not a substitute for maintenance, which includes the preventive
replacement of wearing parts as specified by the manufacturer.
14.3.2.1
Performing the safety checks
1. Check that the associated instructions for use are available in the correct
version.
2. Perform a functional test of the following functions according to the instructions
for use:
– Emergency O
2
delivery
3. Check that the product is in good condition:
– All labels are complete and legible
– No visible damage to the following components:
– Trolley and castors
– Housing parts
– Brakes
– Vaporizer mount
– Water trap and water trap mount
– O-rings of the water trap mount
– Sample line
– Screen
– Gas inlets
– Status display
– Breathing system
– Piston drive and piston diaphragm
– AGS and AGS valves
– Hoses and cables
– Strain relief for compressed gas hoses and cables
– Option: Support arms
– Option: Pressure reducers and their sensor cables
– Fuses that are accessible from the outside are in compliance with the
specified values.
– Check that the country-specific labeling of the gas type matches the screen
display.
– Check the end stops of the support arms for patient monitors.
4. Check the electrical safety in accordance with the IEC 62353 standard.