Troubleshooting
!! 255
Apnea (no pressure)
No breathing or ventilation.
Start manual ventilation!
Check ventilation settings.
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
Check fresh-gas settings and
position of breathing bag.
The breathing hoses are
blocked or leaking.
Check breathing circuit and
tube.
!! 90
Apnea Ventilation
No inspiratory effort of the
patient detected.
Check spontaneous breath-
ing ability of the patient.
Adjust the setting for "Trig-
ger".
Change to pressure-con-
trolled or volume-controlled
ventilation mode.
Use "ALARM RESET" to
downgrade alarm priority.
!
0
Apnea Ventilation
No inspiratory effort of the
patient detected.
Check spontaneous breath-
ing ability of the patient.
Adjust the setting for "Trig-
ger".
Change to pressure-con-
trolled or volume-controlled
ventilation mode.
!!! 100
Backup manual mode acti-
vated
The backup manual switch
has been activated.
Start manual ventilation!
Check vaporizer and fresh-
gas settings.
To exit backup manual mode,
return the switch to its normal
position.
!
100
Backup manual mode acti-
vated
The backup manual switch
has been activated.
Start manual ventilation!
Check vaporizer and fresh-
gas settings.
To exit backup manual mode,
return the switch to its normal
position.
!! 100
Backup speaker failure
The backup speaker for
alarm tones is faulty.
Call Dräger.
Use "ALARM RESET" to
acknowledge alarm.
!
180
Battery charge low
The battery charge is low
and the mains power supply
is not available.
Restore mains power supply.
The breathing system
warmer has been switched
off. Check the breathing cir-
cuit for condensate. Increase
the fresh-gas flow if neces-
sary.
Priority
Alarm
Cause
Remedy