Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
261
Reprocessing
– Expiratory port
– Inspiratory valve (yellow)
– Expiratory valve (blue)
– Fresh-gas decoupling valve (black)
– Piston diaphragm
– Breathing bag arm (option)
– Absorber container and absorber insert
– CLIC adapter (option) with cleaning plate fitted
Drying
25. Allow all the components to dry completely.
Steam sterilization
26. If necessary, steam sterilize the following components:
– Upper part of the breathing system
– Lower part of the breathing system
– Sensor cap of the O
2
sensor
– APL valve
– Inspiratory port
– Expiratory port
– Inspiratory valve (yellow)
– Expiratory valve (blue)
– Fresh-gas decoupling valve (black)
– Piston diaphragm
– Breathing bag arm (option)
– Absorber container and absorber insert
Preparation for fitting and assembly
27. Inspect the following components for damage and
wear:
– Upper part of the breathing system
– Lower part of the breathing system
– Sensor cap of the O
2
sensor
– Incident flow mesh in the inspiratory limb of the
lower part of the breathing system
– APL valve
– Inspiratory port
– Expiratory port
– Incident flow mesh in the expiratory port
– Inspiratory valve (yellow)
– Expiratory valve (blue)
– Fresh-gas decoupling valve (black)
– Piston diaphragm
– Flow sensors
– Seals and sealing rings
28. The following parts must be free from residues:
– Incident flow meshes for the flow sensors
– Valve plates of the non-return valves
Assembling the breathing system
29. Insert the non-return valves:
– Inspiratory valve (yellow)
– Expiratory valve (blue)
30. Insert the fresh-gas decoupling valve (black) and lock it
by turning.
31. Fit the upper part of the breathing system.
32. Fit the APL valve.
33. Fit the flow sensors and ports.
a. Insert the expiratory flow sensor.
b. Screw the expiratory port tight.
c. Insert the inspiratory flow sensor.
d. Screw the inspiratory port tight.
34. CLIC adapter (option)
a. Remove the cleaning plate.
b. Screw on the adapter.
Inserting the piston diaphragm
35. Insert the piston diaphragm.
Inserting the breathing system
36. Insert the breathing system.
Assembling and inserting the O
2
sensor
37. Insert the sensor cap.
38. Fit the O
2
sensor.
Fitting the breathing bag arm (option)
39. Fit the breathing bag arm (option).
Fitting the CO
2
absorber
40. Reusable CO
2
absorber
a. Fit the absorber insert.
b. Fill the reusable CO
2
absorber.
c. Insert the disposable dust filter.
d. Screw the absorber tight.
41. CLIC absorber (disposable absorber)
a. Fit the CLIC absorber.
Fitting patient-specific accessories and consumables
42. Complete the device with the following accessories:
– Breathing hoses
– Breathing bag
– Filter
– Water trap
– Sample line
Checking the operational readiness
43. Connect the device to the mains power supply and turn
it on.
44. Perform a system test.