Getting started
For further information about the configuration, see page 166.
6.4
Checking the operational readiness
Using colors, the
Standby
page (1) indicates whether the system test was
successful and the device is ready for operation.
If the device is not fully operational, the most important irregularities (5) are
displayed along with a recommendation to perform a specific test (3).
Additionally, the current system leakage is displayed in area (5).
To view details regarding the status of the device, touch the
Details...
button (2) or
the
Tests...
button (4), see page 126.
Dräger recommends performing the system test every 24 hours. Otherwise, it will
not be possible to ensure that the device is functional.
3503
4
Color
Meaning
Green
System is fully operational.
Yellow
System is operational with limitations.
There are functional restrictions.
Take further measures to ensure patient safety (e.g., external
monitoring).
Red
System is not operational.