Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
65
Assembly and preparation
4.6
Connecting to the gas scavenging system
The device is equipped with an active or a passive anesthetic gas receiving system
(AGS).
The active anesthetic gas receiving system can be combined with a control valve or
an ejector. Follow the relevant assembly instructions.
►
Make sure that the ventilation slots on the underside of the AGS are not blocked.
4.6.1
Active anesthetic gas scavenging
1.
CAUTION
Risk of ambient air contamination
If the anesthetic gas receiving system is not connected to the disposal system,
contamination of the ambient air with anesthetic gas may result.
►
Connect the anesthetic gas receiving system correctly to the disposal
system.
Connect the scavenging hose (1) to the nozzle on the AGS.
2. Secure the scavenging hose with the clip (2).
3. Connect the gas probe of the scavenging hose to the terminal unit (3) of the
disposal system.
36
4
9
4
36495