Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
227
Reprocessing
Dräger states that oxygen-releasing agents and chlorine-releasing agents may
cause color change in some materials. Color change does not indicate that the
product is not functioning correctly.
13.6 Preparation
13.6.1
Disassembling the patient-specific accessories
1.
WARNING
Risk of infection
Used sample lines and water traps may be infectious due to the breathing gases
that passed through them.
►
Replace the sample line and the water trap regularly in the following
situations:
– If the sample line is connected to the filter on the Y-piece, replace it daily.
– If there is no filter fitted to the Y-piece and the sample line is connected
directly to the Y-piece, replace the sample line after every patient.
►
Remove the sample line from the water trap.
►
Initially leave the water trap fitted to prevent infectious fluid from spurting out.
Remove the water trap only after surface disinfection.
►
Replace the water trap after each patient.
Quaternary ammonium
compounds
Mikrozid sensitive liquid
1)
Schülke & Mayr
Mikrozid sensitive wipes
Mikrozid alcohol-free liquid
Mikrozid alcohol-free wipes
acryl-des
Cleanisept Wipes Maxi
Dr. Schumacher
Surfa'Safe Premium
ANIOS Laboratories
Wip'Anios Excel
Tuffie 5
Vernacare
1) Virucidal against enveloped viruses
Class of active ingre-
dient
Surface disinfectants
Manufacturer