Troubleshooting
!
255
Inspiratory halothane low
The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakage.
The soda lime has dried out. Replace the soda lime.
!!! 255
Inspiratory isoflurane high Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Check vaporizer and fresh-
gas settings.
!! 255
Inspiratory isoflurane high Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Check vaporizer and fresh-
gas settings.
!
255
Inspiratory isoflurane low
The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakage.
The soda lime has dried out. Replace the soda lime.
!! 10
Inspiratory N
2
O high
Inspiratory N
2
O exceeds 82
%.
Check fresh-gas composi-
tion.
Press the O
2
+ button to flush
the breathing system.
!
255
Inspiratory O
2
measure-
ment temporarily inaccu-
rate
The O
2
measurement accu-
racy is currently reduced.
Perform the system test.
!!! 255
Inspiratory sevoflurane
high
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Check vaporizer and fresh-
gas settings.
!! 255
Inspiratory sevoflurane
high
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Check vaporizer and fresh-
gas settings.
!
255
Inspiratory sevoflurane low The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakage.
The soda lime has dried out. Replace the soda lime.
!!! 255
Inspiratory xMAC high
The inspiratory anesthetic
gas concentration has
exceeded 3 xMAC for more
than 30 seconds. The expira-
tory anesthetic gas concen-
tration has exceeded 2.5
xMAC for more than 30 sec-
onds.
Check vaporizer and fresh-
gas settings.
The inspiratory anesthetic
gas concentration has
exceeded 5 xMAC.
Check vaporizer and fresh-
gas settings.
Priority
Alarm
Cause
Remedy