Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
81
Operating concept
5.1.3 Therapy
bar
The following illustration shows the expanded therapy bar for the electronically
controlled gas mixer:
Start values
Arrows
on the scales of the therapy controls mark the start values resulting from
the patient data and start settings. The start values can be configured, see
page 166.
Linked therapy controls
Certain parameters can be linked to other parameters. If one parameter is changed,
the linked parameter is also selected and changed. Among other things, this applies
to the adjustment of ventilation pressures, ventilation times or during electronically
controlled fresh-gas delivery.
Example: The device can be configured so that a change to the
PEEP
setting
automatically causes a change to
Pinsp
; as a result, the difference between
PEEP
and
Pinsp
and therefore the tidal volume remain constant.
Linking therapy controls, see page 175.
34752
No.
Designation
1
Name of active ventilation mode
2
Tabs
3
Notification field
4
Buttons to expand and collapse the therapy bar
5
Field with additional information:
– More values
– Spontaneous breathing activity of the patient
6
Therapy controls for ventilation parameters and buttons for synchronizing
the breaths
7
With electronically controlled gas mixer:
Therapy controls and buttons for fresh-gas delivery
2
3
4
1
5
4
6
7