Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
107
Operation
An insufficient fresh-gas flow may trigger the
Fresh gas low or leakage
or
Emergency air inlet activated
alarms, for example.
7.6
Customizing the screen display
7.6.1 Available
views
The following views are available:
Standard view
Up to three waveforms (1), (2), and (3) are displayed along with their associated
parameter fields.
Expert view
In addition to the standard view, the 4 additional parameter fields (4), (5), (6), and
(7) are displayed.
Too low
Red
The fresh-gas flow is too low.
Check the fresh-gas flow.
Check the position of the breathing
bag.
Refill bag
Red
The fresh-gas flow is too low.
Check the filling of the breathing bag. If
necessary, fill up the breathing bag,
e.g., with the O
2
flush.
Evaluation
Color
Meaning