Troubleshooting
!!! 120
Ventilator failure
The piston ventilator or at
least one pressure sensor is
faulty.
Start manual ventilation!
When the case has been
completed, perform a sys-
tem test.
If the problem persists, call
Dräger.
!
100
Water trap disconnected? Infinity ID water trap is not
correctly connected.
Check water trap.
!
100
Water trap expired
Accessory has been used
too long.
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
acknowledge alarm.
!
170
Water trap full
The water trap of the gas
measurement is full.
Check water trap.
Sample line is occluded.
Check sample line, water
trap, and patient-side filter.
!!! 255
Wrong patient connection? The patient is connected to
the internal breathing sys-
tem, but there is no oxygen
delivery.
Change to Man/Spon mode
and ventilate the patient
manually.
Breathing activity has been
detected at an inactive
breathing system.
Check if the patient is con-
nected to the correct breath-
ing system.
!! 80
xMAC low
Inspiratory and expiratory
gas concentrations are lower
than the automatically calcu-
lated limit.
Check patient condition.
Check filling level. Refill if
necessary.
Check vaporizer setting.
Check the breathing system
and the breathing bag for
leakages.
If the current xMAC is
acceptable, use "ALARM
RESET" to acknowledge the
alarm.
!
170
xMAC low
Inspiratory and expiratory
gas concentrations are lower
than the automatically calcu-
lated limit.
Check patient condition.
Check filling level. Refill if
necessary.
Check vaporizer setting.
Check the breathing system
and the breathing bag for
leakages.
If the current xMAC is
acceptable, use "ALARM
RESET" to acknowledge the
alarm.
Priority
Alarm
Cause
Remedy