280
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Time until measured CO
2
values are displayed
(with at least reduced accuracy)
<90 s
Typical delays
<5 s
Cross sensitivity
None with respect to alcohol (<3000 ppm in
blood),
Acetone (<1000 ppm), methane, water vapor,
NO and CO
Drift
Compensated by automatic cyclic calibration
During the automatic calibration, ambient air is
fed to the breathing system and used as the
sample gas.
O
2
Range
0 to 100 Vol%
Accuracy
±(2.5 Vol% + 2.5 % relative)
Resolution of displayed value
1 Vol%
T
10...90
<500 ms
CO
2
Range
0 to 13.6 Vol%
0 to 13.6 kPa
0 to 102 mmHg
Accuracy
±(0.43 Vol% + 8 % relative)
±(0.43 kPa + 8 % relative)
±(3.3 mmHg + 8 % relative)
Resolution of displayed value
0.1 Vol%
0.1 kPa
1 mmHg
T
10...90
<350 ms
N
2
O
Range
0 to 100 Vol%
Accuracy
±(2 Vol% + 8 % relative)
Resolution of displayed value
1 Vol%
T
10...90
<500 ms
Anesthetic gases
Range
Halothane
0 to 8.5 Vol% (kPa)
Isoflurane
0 to 8.5 Vol% (kPa)
Enflurane
0 to 10 Vol% (kPa)
Sevoflurane
0 to 10 Vol% (kPa)
Desflurane
0 to 20 Vol% (kPa)
16.15
Gas measurement with PGM (continued)