142
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
Press the
O
2
+
key. The inflow of gas is audible. Keep the key pressed until the
pressure exceeds 15 hPa (cmH
2
O).
5. Pull off the Y-piece and plug it on again. The pressure drops.
8.4.3.1 Special
test
settings
1. Test breathing circuit
Prerequisite:
– Patient-gas measurement module present
– Sample line is connected during test: On
– Electronically controlled gas mixer or mechanically controlled gas mixer with
O
2
detection
A check is made to ensure that the inspiratory hose, expiratory hose, and
breathing bag hose are connected correctly.
This can be deactivated in the System setup > System > System test dialog.
2. Verify O
2
delivery
Prerequisite:
– Patient-gas measurement module present
– Sample line is connected during test: On
– Electronically controlled gas mixer or mechanically controlled gas mixer with
O
2
detection
A check is made to ensure that the O
2
supply is actually delivering O
2
. This is
checked at the Y-piece, using the connected sample line.