![Dräger Atlan A300 Instructions For Use Manual Download Page 182](http://html.mh-extra.com/html/drager/atlan-a300/atlan-a300_instructions-for-use-manual_2526728182.webp)
182
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
Prerequisite: USB mass storage device has been connected to the USB port.
►
Open the
System setup
>
Im/Export config.
>
Im/Export config.
page (1).
The configurations saved on the USB mass storage device are displayed in a
list (4). If not all of the configurations are displayed, delete all configurations from
the USB mass storage device that are not needed or move them to a subdirectory
on the USB mass storage device.
The following settings are neither imported nor exported:
–
Device name
–
Date and time
–
IP address
11.4.1 Importing
the
configuration
1. Touch one of the configurations in the list (4).
2. Touch the
Import
button (2) and confirm.
3. Restart the device.
4. Check the device configuration for correctness.
11.4.2
Exporting the configuration
►
To export configurations, touch the
Export
button (3) and confirm.
11.5
Activating software options
The following software options require an activation code to be entered, followed by
activation:
– Spontaneous breathing support
– AutoFlow
– Advanced trends
– Advanced ventilation monitoring
– Advanced gas monitoring
36030
4
1
3
2
1
System setup