Principles of operation
17.3
Description of the ventilation drive
17.3.1 Principle
of
operation
The device has a piston drive.
The piston drive is connected directly to the breathing system. The electric
motor (3) moves the piston (4) by means of a spindle (2). Thus, the fresh gas is
moved out of the piston diaphragm and through the breathing system to the patient.
The user can check through a viewing window whether the piston is moving. During
inspiration, the ventilator applies the tidal volume with the required pressure and at
a defined respiratory rate.
4016
2
No.
Designation
1
Piston diaphragm
2 Spindle
3
Electric motor
4
Piston