Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
3.
WARNING
Risk of hypoventilation and incorrect gas measurement
Reuse of the disposable dust filter can increase the filter resistance and impair
the ventilation function of the device.
►
If soda lime from third-party manufacturers is used, e.g., granular soda lime,
use a disposable dust filter and replace it with every change of the soda lime.
If soda lime from third-party manufacturers is used, insert a new disposable dust
filter (3). Only use dust filters from the list of accessories.
Only use undamaged filters.
4. Attach the CO
2
absorber (4) to the breathing system from below. Rotate it in the
direction of the arrow until it reaches the stop.
Follow the instructions for use for the particular soda lime.
13.12
Preparation before the next use
13.12.1 Fitting
patient-specific accessories and consumables
►
WARNING
Risk due to particles and dust
In order to protect the patient from particles and dust, a filter must be used
between the inspiratory limb of the breathing system and the patient.
►
Use a filter at the Y-piece or at the inspiratory port.
Complete the device with the following accessories:
– Breathing hoses
– Breathing bag
– Filter
– Water trap
– Sample line
Observe the following information: "Selecting and connecting patient-specific
accessories", page 67.
39515