Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
257
Reprocessing
5. Put the breathing system cover (5) in place and click it into position.
13.11.6 Fitting
the
breathing bag arm (option)
1. Plug the breathing bag arm (1) onto the socket (2) on the breathing system.
Tighten the two knurled screws. Check that the breathing bag arm is held
securely.
2. Align the breathing bag arm so that collisions with other components are
prevented.
13.11.7
Fitting the reusable CO
2
absorber
If there is no CLIC absorber being used, use the reusable CO
2
absorber.
1. Push the absorber insert (1) into the absorber container (2).
2.
CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
►
Handle the soda lime carefully and do not spill it.
Fill the CO
2
absorber with fresh soda lime to the upper mark.
Recommendation: Use Drägersorb 800 Plus or Drägersorb Free.
36505
36506
39516
1
2