Tests
The test progress is displayed in area (8) while the automatic test is running. All test
results are displayed in area (9).
After the test, the final test result is displayed on the standby screen, see page 87.
8.3.1.1
Test interruption due to irregularities
If an irregularity is detected during the automatic test, the following occurs:
– The test is interrupted.
– An acoustic signal sounds. This signal is repeated every 15 seconds.
– Information on the cause and remedy are displayed.
Remedying the cause:
1. Remedy the cause of the interruption.
2. Touch the
Repeat
button (1) and repeat the test of the component.
Accepting the irregularity:
►
Touch the
Accept
button (2) and continue the test.
Accepted irregularities prevent the total result of the system test from indicating
"fully operational" and are protocoled in the logbook.
8.3.2
Manual check for leakage
8.3.2.1
Checking a vaporizer for leakage
The leakage of a vaporizer can be determined with 2 tests:
– Quick tightness check: Checks whether there is leakage.
– Complete leakage test: Determines the leakage value.
Prerequisites:
– The leakage test of the device has been performed and passed.
– The vaporizer is mounted directly and securely on the plug-in adapter.
– The filling inlet is closed.
– The vaporizer is closed. The control dial is in the
0
position.
– The breathing circuit is correctly connected.
– The APL valve is set to 30.
– The Y-piece is sealed.
35088
1
2
Test details