Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
153
Alarms
10.2.1
Optical alarm signals
In the event of an alarm, the device displays the relevant alarm message in the
alarm message field. For certain alarms, the parameter field of the parameter
triggering the alarm will flash.
In the alarm message field (2), up to 8 alarms can be displayed at a time. If more
alarms occur, the
All alarms
button (1) is displayed. Touching this button opens the
Alarms
>
Current alarms
dialog with information about all active alarms, see
page 154.
10.2.2
Acoustic alarm signals
It always is the alarm with the highest priority that is acoustically signaled. The
signal is emitted until either the cause of the alarm is remedied or the alarm silence
key is pressed.
In situations where several alarms occur at the same time, alarms with higher
priority may only be signaled with a 5-tone sequence instead of a 10-tone
sequence.
Regardless of the set alarm volume, the
No O
2
delivery
alarm is issued at
maximum volume.