270
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Information
The maximum flow specified under STPD con-
ditions can drop linearly from 15 L/min at
750 hPa ambient pressure to 13.5 L/min at
620 hPa ambient pressure. At 620 hPa ambient
pressure, an STPD flow of 13.5 L/min corre-
sponds to an ATPD flow of 22 L/min.
Accuracy
±10 % or ±50 mL/min (the larger value applies)
16.6 Fresh-gas
delivery
with
mechanically controlled gas
mixer
Standardization
All data are standardized to STPD conditions.
O
2
concentration
21 to 100 Vol%
Setting range for fresh-gas flow
0 to at least 12 L/min (O
2
, Air, and N
2
O)
Electronic measurement of fresh-gas flow
Range
0 to 15 L/min (O
2
, Air, and N
2
O)
Accuracy
±10 % or ±0.12 L/min (the larger value applies)
Resolution of the value displayed on the screen
0.01 L/min (from 0 to 0.2 L/min)
0.02 L/min (from 0.2 to 0.5 L/min)
0.05 L/min (from 0.5 to 1 L/min)
0.10 L/min (from 1 to 15 L/min)
Resolution of the value displayed in the status
display
0.1 L/min
Total flow tube
Standardization
All data are standardized to STPD conditions.
Range
0 to 10 L/min
Accuracy
±10 % of the set value at 100 % O
2
and for
flows >1 L/min
Resolution of displayed value
0.5 L/min
16.7 Ventilator
Design
Electronically driven ventilator, fresh-gas
decoupled
Time-based settings
Respiratory rate RR
3 to 100 /min
±10 % of the set value or ±1 /min (the larger
value applies)
16.5
Fresh-gas delivery with electronically controlled gas
mixer (continued)