Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
55
Assembly and preparation
4.3
Before first operation
The breathing system must be reprocessed before the device is operated for the
first time.
4.3.1 Inserting
the
O
2
sensor cell
If the device is equipped with an O
2
sensor, an O
2
sensor cell must be inserted into
the O
2
sensor.
1. Disassemble the O
2
sensor (see "Disassembling the O
2. Insert the new O
2
sensor cell (1) into the O
2
sensor.
After the O
2
sensor cell has been inserted, the initialization phase of the O
2
sensor takes place. The initialization lasts 30 minutes. The 30 minutes start when
the mains power supply is established.
4.3.2
Assembling and inserting the O
2
sensor
1. Place the sensor cap (1) on the O
2
sensor.
39637
1
37178