![Dräger Atlan A300 Instructions For Use Manual Download Page 160](http://html.mh-extra.com/html/drager/atlan-a300/atlan-a300_instructions-for-use-manual_2526728160.webp)
160
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Alarms
10.4.1
Activating the alarms related to volume
The upper alarm limit for
MV
is deactivated at the start of the ventilation, but can be
set during the ventilation.
The
MV low
alarm is delayed in certain cases and is issued as follows:
– No sooner than 45 seconds after a case starts,
– No sooner than 45 seconds after changing to a mode with greater respiratory
support, see page 299,
– No sooner than 45 seconds after an
Apnea (no flow)
alarm or an
Apnea
(no pressure)
alarm.
10.4.2
Resetting the Apnea (no CO
2
) alarm
Prerequisite: The device has the "Integrated patient-gas measurement module"
option.
When changing to a ventilation mode with higher respiratory support, the
Apnea
(no CO
2
)
alarm is reset. If the apnea situation persists, an alarm appears after the
time specified in table "Alarm delay and alarm escalation".
10.5
Alarm delay and alarm escalation
To prevent unnecessary alarms, some alarms are not displayed immediately after a
limit violation, but after a delay. In addition, certain circumstances can cause the
alarm priority to change.
Gas measurement alarms with integrated patient-gas measurement module
Alarm
Priority
Priority
Priority
Note
(Low)
Caution
(Medium)
Warning
(High)
inCO
2
high
etCO
2
high
etCO
2
low
FiO
2
high
Inspiratory N
2
O high
---
After 2 consecutive
respiratory phases and
15 seconds
---
inAgent low
After 2 consecutive
respiratory phases and
15 seconds
---
---
FiO
2
low
---
---
After 2 consecutive
respiratory phases and
15 seconds
or
after 30 seconds if no
respiratory phases are
detected